ISO 13004:2022

Sterilization of health care products — Radiation — Substantiation of selected sterilization dose: Method VDmaxSD
Autor: ISO
Código ICS: 11.080.01
Vigente
$290.360

Alcance

This document describes a method for substantiating a selected sterilization dose of 17,5 kGy, 20 kGy, 22,5 kGy, 27,5 kGy, 30 kGy, 32,5 kGy or 35 kGy that achieves a sterility assurance level (SAL) of 10-6 or less for radiation sterilization of health care products. This document also specifies a method of sterilization dose audit used to demonstrate the continued effectiveness of the substantiated sterilization dose.NOTE 1   Selection and substantiation of the sterilization dose is used to meet the requirements for establishing the sterilization dose within process definition in ISO 11137-1.This document does not apply to other sterilization doses than the substantiation of a selected sterilization dose of 17,5 kGy, 20 kGy, 22,5 kGy, 27,5 kGy, 30 kGy, 32,5 kGy or 35 kGy. The method is not used for the substantiation of a selected sterilization dose if the average bioburden of the entire product item exceeds the limit specified for the selected sterilization dose (see Table 3).NOTE 2   The methods for substantiation of selected sterilization doses of 25 kGy and 15 kGy are not included in this document. They are described in ISO 11137-2.If the decision is made to use this method of sterilization dose establishment, the method is intended to be followed in accordance with the requirements (shall) and guidance (should) stipulated herein.

Especificaciones de la Norma
Fecha de Publicación 07/10/2022
Título Secundario Stérilisation des produits de santé — Irradiation — Justification de la dose stérilisante choisie: Méthode DVmaxDS
Páginas Técnicas 54
Idioma Inglés